Soft matter
Sample environment equipment
Reometer
・Anton-Paar GmbH MCR 302
-
A double-cylindrical quartz glass container is used.
-
Gap space between two cylinders:1 mm
-
It is available for the ReoSANS along both radial and tangential directions
-
Temperature range:-50~150℃
-
Torque:1 nNm ~ 200 mNm
-
Angular velocity:10-9 ~ 314 rad/s
-
Available instrument:BL15
-
It is available for the off-line experiment in the laboratory out of the MLF.
.
High-precesion gas/ vapor adsorption analyzer
・MicrotracBEL BELSORP-max
-
A quartz glass container is used.
-
Inner diameter:φ19 mm
-
Sample thickness:1 or 2 mm
-
Observable pore size distribution:3.5 ~ 5,000 Å
-
Adsorbate:N2, Ar, H2O, D2O etc.
-
Available instrument:BL15
-
It is available for the off-line experiment in the laboratory out of the MLF.
Members
Shin-ichi Takata(JAEA, J-PARC Center)
Toshiaki Morikawa(CROSS)