Top » Sample environment » High-magnetic field

High-magnetic field

Superconducting magnet

7T superconducting magnet (vertical magnetic field)

Production company: Scientific MAGNETICS

Maximum magnetic field:± 6.85 T

Bore diameter:50 mm

Temperature range:3~300 K(attainable lowest temperature:1.5 K@VTI、Single shot:1.07 K@VTI)

Available scattering angle range:±10°(vertical)、330°(horizontal)

Attachment size:[Vacuum chamber] JIS800 flange, [Sample stage] 700 mm in diameter

・Aluminum window

・ω rotation(±180°)

・Usually, liquid He is added every two days.

・It takes about 2 days to prepare the magnet for the experiment.

・The dilution refrigerator insert is available.

Used instruments:BL01(max. 3.5 T), BL14(max. 4 T), BL15. BL17, BL18(max. 6.8 T)

The maximum magnetic field is determined by the magnetic environment on the beam line.

.

Pulsed magnet

Pulsed power supply with a capacitor bank

Assembled in J-PARC

Output current:8 kA(maximum)

Pulse-width:3-6 ms

Repetition rate:one pulse per 10 seconds(fastest)

Charge voltage:2 kV(maximum)

Charge energy:16 kJ

It is transportable.

Used instrument:BL10、BL12

Reference: [1]

Cryostat for the pulsed magnet

Maximum magnetic field:30 T(maximum)

Sample temperature:3 K (minimum)

Repetition rate:one pulse per 1-10 min

Scattering angle:42°(maximum)

Inductance of the coil:670 μH(@100 Hz)

Size of the coil:Outer diameter: 44 mm, Inner diameter: 14mm, width: 21mm

Cryogenic system:GM refrigerator(HE05RK、ULVAC)

Sample temperature is variable by using a heater.

Used instrument:BL10

If you would like to use the pulsed magnet system, please consult to the MLF SE team.

Reference [1]

.

Members

<<< Suerconducting magnet >>>

Ryuta Takahashi(JAEA, J-PARC Center)

Motoyuki Ishikado(CROSS)

Yasuhiro Yamauchi(JAEA, J-PARC Center)

Sora Ishimaru(JAEA, J-PARC Center)

Hirotsugu Morimizu(JAEA, J-PARC Center)

Seiko Ohira-Kawamura(JAEA, J-PARC Center)

<<< Pulsed magnet >>>

Masao Watanabe(JAEA, J-PARC Center)

.

References

[1] M. Watanabe and H. Nojiri, "Pulsed magnet system at MLF in J-PARC", Journal of Neutron Research, 21 (2019) 39.