High-magnetic field
Superconducting magnet
7T superconducting magnet (vertical magnetic field)
Production company: Scientific MAGNETICS
Maximum magnetic field:± 6.85 T
Bore diameter:50 mm
Temperature range:3~300 K(attainable lowest temperature:1.5 K@VTI、Single shot:1.07 K@VTI)
Available scattering angle range:±10°(vertical)、330°(horizontal)
Attachment size:[Vacuum chamber] JIS800 flange, [Sample stage] 700 mm in diameter
・Aluminum window
・ω rotation(±180°)
・Usually, liquid He is added every two days.
・It takes about 2 days to prepare the magnet for the experiment.
・The dilution refrigerator insert is available.
Used instruments:BL01(max. 3.5 T), BL14(max. 4 T), BL15. BL17, BL18(max. 6.8 T)
The maximum magnetic field is determined by the magnetic environment on the beam line.
.
Pulsed magnet
Pulsed power supply with a capacitor bank
Assembled in J-PARC
Output current:8 kA(maximum)
Pulse-width:3-6 ms
Repetition rate:one pulse per 10 seconds(fastest)
Charge voltage:2 kV(maximum)
Charge energy:16 kJ
It is transportable.
Used instrument:BL10、BL12
Reference: [1]
Cryostat for the pulsed magnet
Maximum magnetic field:30 T(maximum)
Sample temperature:3 K (minimum)
Repetition rate:one pulse per 1-10 min
Scattering angle:42°(maximum)
Inductance of the coil:670 μH(@100 Hz)
Size of the coil:Outer diameter: 44 mm, Inner diameter: 14mm, width: 21mm
Cryogenic system:GM refrigerator(HE05RK、ULVAC)
Sample temperature is variable by using a heater.
Used instrument:BL10
If you would like to use the pulsed magnet system, please consult to the MLF SE team.
Reference [1]
.
Members
<<< Suerconducting magnet >>>
Ryuta Takahashi(JAEA, J-PARC Center)
Motoyuki Ishikado(CROSS)
Yasuhiro Yamauchi(JAEA, J-PARC Center)
Sora Ishimaru(JAEA, J-PARC Center)
Hirotsugu Morimizu(JAEA, J-PARC Center)
Seiko Ohira-Kawamura(JAEA, J-PARC Center)
<<< Pulsed magnet >>>
Masao Watanabe(JAEA, J-PARC Center)
.
References
[1] M. Watanabe and H. Nojiri, "Pulsed magnet system at MLF in J-PARC", Journal of Neutron Research, 21 (2019) 39.